0207-050-702 Rev-L
Features
F
H
G
E
K
J
I
E
Applied Part(s) – The tip assembly and handpiece body are
A
the applied parts as defined by the manufacturer according to
the standards listed under Product Safety Certification in the
Specifications section.
B
Latch Mechanism – The mechanism is used to release the tip
assembly from the handpiece.
C
Trigger – An activated trigger provides irrigation to the surgical/wound
site.
D
Trigger Lock – The lock is used to hold the trigger at either the high
irrigation flow or low irrigation flow setting.
E
Irrigation Tubing/Power Cord Combination – The irrigation tubing
provides a fluid pathway from the irrigation bag to the handpiece tip
assembly. The power cord conducts electrical energy from the battery
power pack to the handpiece.
F
Suction Clamp – The clamp is used to open and close the suction
tubing pathway to control suction at the handpiece tip assembly (not
included with all handpiece configurations).
G
Suction Tubing – The suction tubing provides a suction pathway
from the handpiece tip assembly to the vacuum canister connected
to the facility vacuum source (not included with all handpiece
configurations).
H
Suction Tube Port – The port allows connection of the suction tubing
to the handpiece (not shown).
I
Battery Power Pack (removable batteries) – The battery pack
provides power to the handpiece and allows easy removal of internal
batteries. (REF 0207-580-000, REF 0207-583-000, REF 0207-585-
000, REF 0207-586-000, REF 0207-589-000 and REF 0207-590-000
only)
www.stryker.com
Print Date: Aug 29, 2016 01:59:30 PM
0000004063, Rev. L Effective Date: -
A
B
C
D
K
J
L
E
Features (continued)
J
Irrigation Tube Spike – The spike is used to pierce and connect the
handpiece irrigation tubing to the irrigation bag.
K
Irrigation Tube Spike Cap – The protective spike cap is installed over
the sharp irrigation tube spike to maintain sterility and prevent damage
or injury.
L
Battery Power Pack (sealed) – The battery pack provides power to
the handpiece.
M
Suction Tube Adapter – The adapter is used for a large diameter
port on a vacuum fluid collection canister and may be removed if
necessary.
Definitions
The symbols located on the equipment and/or labeling are defined in this section
or in the Symbol Definition Chart. See the Symbol Definition Chart supplied with
the equipment.
SYMBOL
M
G
Instructions
To Assemble the Handpiece
WARNINGS:
• ALWAYS use aseptic technique when handling the sterile handpiece and
irrigation tubing.
• DO NOT reuse the handpiece for surgical applications or applications
requiring a sterile field.
• ALWAYS use aseptic technique when removing the irrigation tube spike cap.
Failure to comply may cause the irrigation tube spike to become contaminated
before insertion into the irrigation bag.
NOTES:
• Make sure a standard IV pole is equipped with an irrigation bag.
• Make sure the top of the fluid contained in the irrigation bag is less than 36
inches above the handpiece at all times after connecting the handpiece.
DEFINITION
Consult Instructions for Use
General warning sign
Battery lid removal tabs
(REF 0207-580-000, REF 0207-583-000, REF 0207-
585-000, REF 0207-586-000, REF 0207-589-000 and
REF 0207-590-000 only)
Trigger Lock
Type BF Applied Part
Sterilized using irradiation
Direct Current (DC)
Do not re-use
Do not use if package is damaged
EN
3