8. ERROR MESSAGES

If an error occurs during the measurement, the measurement is interrupted and an error
message, e.g. «ERR 3», is displayed.
Error  
 
«ERR  1»    
 
«ERR  2»  
 
«ERR  3»  
«ERR  5»  
«HI»  
«LO»    
 
Note:
 Note:
- When an error occurred during the measurement, the reading is not stored in the memory.
-
When  an  error  occurred  during  the  measurement,  the  reading  is  not  stored  in  the  memory.  
 
 
9.    
Safety,  Care,  Accuracy  Test  and  Disposal    
 
9.1     Safety  and  protection  
9.1.1  
Follow  instructions  for  use.  This  document  provides  important  product  operation  and  safety  
 
information  regarding  this  device.  Please  read  this  document  thoroughly  before  using  the  
 
device  and  keep  for  future  reference.  
9.1.2  
This  device  may  only  be  used  for  the  purposes  described  in  these  instructions.  The  
 
manufacturer  cannot  be  held  liable  for  damage  caused  by  incorrect  application.  
9.1.3  
This  device  comprises  sensitive  components  and  must  be  treated  with  caution.  Observe  the  
 
storage  and  operating  conditions  described  in  the  «Technical  Specifications»  section.  
9.1.4  
Protect  it  from:  
 
-­  water  and  moisture  
RBP-100
 
-­  extreme  temperatures  
 
-­  impact  and  dropping  
 
-­  contamination  and  dust  
 
-­  direct  sunlight  
 
-­  heat  and  cold  
9.1.5  
The  cuffs  are  sensitive  and  must  be  handled  with  care.  
Description  
Potential  cause  and  remedy  
Signal  to  weak  
The  pulse  signals  on  the  cuff  are  too  weak.  Re-­position  
the  cuff  and  repeat  the  measurement.  
Error  signal    
During  the  measurement,  error  signals  were  detected  by  
the  cuff,  caused  for  instance  by  movement  or  muscle  
tension.  Repeat  the  measurement,  pay  attention  the  
patient  is  keeping  his  arm  still.  
Abnormal  cuff  
An  adequate  pressure  cannot  be  generated  in  the  cuff.  
pressure  /  inflation  /  
A  leak  may  have  occurred.  Check  that  the  cuff  is  
deflation  
correctly  connected  and  is  not  too  loose.  Replace  the  
batteries  if  necessary.  Repeat  the  measurement.  
Abnormal  result  
The  measuring  signals  are  abnormal  and  no  result  can  
therefore  be  displayed.  Read  through  the  checklist  for  
performing  reliable  measurements  and  then  repeat  the  
measurement.  
Pulse  or  cuff  Pressure  
The  pressure  in  the  cuff  is  too  high  (over  299  mmHg)  
too  high  
OR  the  pulse  is  too  high  (over  200  beats  per  minute).  
The  patient  needs  to  relax  for  5  minutes  and  then  repeat  
the  measurement.  
                           when  pulse  or  pressure  are  lying  above  these    
                           limits.  
Pulse  too  low  
The  pulse  is  too  low  (less  than  40  beats  per  minute).  
Repeat  the  measurement.  
                           when  pulse  is  below  this  limit.    
 Note:
-
Irregular  heartbeat  detection  is  inactivated      
 Note:
-
Irregular  heartbeat  detection  is  inactivated    