9.2 DEVICE CARE

Cleaning and disinfection of medical products are meant to protect patients, users, and
third persons and lead to value retention of medical products. Due to product design and
materials used, there is no possibility to define the maximum limit of re-processing cycles.
Lifetime of medical products is determined by its function and gentle use. Before sending
back defective products for repair, the described re-processing cycles have to be applied
and followed.
Use a soft cloth and follow one of the methods listed for cleaning the exterior of the device:
9.2.1 Mild soap and water
9.2.2 Hydrogen perxodie solution (3% diluted with water)
9.2.3 Sodium hypochlorite solution (1 to 10 dilution of household chlorite bleach in water).
Note:
- To prove if all display segments and the display illumination unit are working correctly,
press and hold the "+" (5) or "-" (6) buttons. All display segments appear and the display
illumination should activate.
-
To  prove  if  all  display  segments  and  the  display  illumination  unit  are  working  correctly,  press  
 
   and  hold  the  "+"  (5)  or  "-­"  (6)  buttons.  All  display  segments  appear  and  the  display    
 
   illumination  should  activate.    
 

9.3 CLEANING THE CUFF

 
9.3     Cleaning  the  cuff    
Carefully  remove  spots  on  the  cuff  with  a  damp  cloth  and  soapsuds.  
Carefully remove spots on the cuff with a damp cloth and soapsuds.
 
 
 
9.4     Accuracy  test    
We  recommend  this  device  is  tested  for  accuracy  every  2  years  or  after  mechanical  impact  (e.g.  being  
dropped).  Please  contact  your  local  Riester  Service  to  arrange  the  test  (see  foreword).  

9.4 ACCURACY TEST

 
Monitoring  of  instruments  
All  countries  except  for  Germany:  
The  respective  legal  provisions  apply  for  all  countries,  except  for  Germany.  The  reference  manometer,  
We recommend that this device is tested for accuracy every 2 years or after mechanical
which  is  used  for  calibration,  must  be  traceable  to  national  and  international  measurement  standards.  
impact (e.g. being dropped). Please contact your local Riester Service to arrange the test.
 It  is  not  allowed  to  make  changes  to  the  device!  
Monitoring of instruments
 
9.4.1.   Verification  of  Calibration  (Static)  
All countries except for Germany:
 
The respective legal provisions apply for all countries, except for Germany. The reference
Test  equipment:  
manometer, which is used for calibration, must be traceable to national and international
 
measurement standards.
Equipment  Required:  
9.4.1.1   Calibrated  reference  manometer  (1)  or  equivalent  (pressure  reference).  
9.4.1.2   500ml  volume  (3).  
It is not allowed to make changes to the device!
9.4.1.3   Hand  Inflation  Bulb  (4)  with  Bleed  valve  (pressure  control).  
9.4.1.4   Blood  pressure  device  (2)  to  be  tested.  
 
To  perform  Calibration  Verification,  proceed  as  follows:  
9.4.1.6   Make  the  necessary  connections  using  the  materials  listed  in  the  Equipment  
 
Required  list.  See  Figure  1  for  pneumatic  connections.  
RBP-100
9.4.1.7   Enter  the  calibration  verification  mode  at  the  RBP-­100: The  device  must  be  switched  off  
 
Press  the  Start  /  Stop  button  and  turn  on  the  power  switch  at  the  same  time.  
 
 
WARNING:  Do  not  wash  the  cuff  in  a  washing  machine  or  dishwasher!    
   