Publicidad

Video Recording
Snapshot
Current Procedure
Image
Adjustments
Archive
Settings
*Access to Archive without log in can be enabled by Administrator in Settings.
**Service User does not have permission to add and change user profiles.
4.2. Preparation and Inspection of the Displaying Unit
1.
Closely examine the displaying unit and other parts for any damage. Do not use the
displaying unit if it is damaged in any way A .
2.
Clean and disinfect the displaying unit (see section 5).
3.
If needed, rotate the displaying unit to have the connector ports for the visualization
device oriented left or right in your preferred direction B .
4.
Choose where to place the displaying unit.
a) Place the displaying unit on a solid flat surface by folding out the stand on the back of
the displaying unit Ca .
b) Mount the supplied bracket on a pole. First, fold the stand on the back upwards. Hold
the displaying unit with two hands and place the bottom edge in the bracket. Push
displaying unit towards the bracket until it clicks in place. Make sure that the bracket and
the displaying unit is securely in place by tightening the bolts on the bracket Cb .
5.
Turn the displaying unit ON by pressing the power button
interface is loading, a live image is available if an Ambu visualization device is connected.
6.
Check the battery indicator
interface appears E . The colour of the battery icon indicates the power level. It is
recommended to charge the displaying unit if the battery icon is orange or red.
3:00
Green = Battery level > 40%
= Charging
Charge the displaying unit if necessary by connecting the power supply to a mains socket
and inserting the power plug into the power inlet on the back of the displaying unit. A
lightning icon
unit is charging. Make sure the power supply is present and working at any time. It is
recommended to locate the nearest power outlet before starting any procedure.
7.
An external monitor can be connected to the HDMI or SDI port located in the back of
the displaying unit. USB ports for file export or software updates are placed on the side
(see section 2.3). Please consult the manual of the external equipment for further
information. F
To avoid risk of electrical shock only connect ancillary equipment, which is approved as
medical electrical equipment.
8.
Connect the Ambu visualization device to the displaying unit by plugging the cable
connector into the corresponding connector port. Make sure that the colours are
matching and the arrows are aligned G .
9.
Verify that a live video image appears on the screen, by pointing the distal end of the
Ambu visualization device towards an object, e.g. the palm of your hand. Make sure that
the live image has the correct orientation H .
10
x
x
x
x
x*
1:45
in the lower left corner of the screen when the user
0:45
Orange =
Battery level < 40%
is shown in the battery indicator on the screen when the displaying
x
x
x
x
x
x
x
x
x
x
x
D . While the user
0:25
Red =
Battery level < 20%
x
x
x
x
x**