Live image on
Processing failure
the displaying
unit screen is
only working
while booting
No image or
External screen is
User Interface
turned OFF
on the external
Issue with cable or
screen
cable connection
Wrong input channel is
selected on the
external monitor
SDI – External screen
does not support
3G-SDI
The colours,
Image settings on
contrast,
external screen are
sharpness and
not correct
brightness look
different on the
external screen
than on the
displaying
unit screen
Force the displaying unit to turn OFF
1.
Press the power button for 10 seconds
to force Turn OFF.
2.
Press the power button again to turn
the displaying unit back on.
If this does not work reset displaying
unit hardware
1.
Remove all connected items including
the power supply
2.
Press the hardware reset button on
the back of the displaying unit*
3.
Reconnect the power supply
4.
Press the power button to turn it
back ON
Turn ON external screen
Check that the SDI or HDMI cable is
connected properly
If possible, try a new cable
Make sure the correct input channel is
selected on the external monitor
Make sure that your external screen
supports 3G-SDI (1920 x 1080p 60 fps.)
Adjust the image settings on the external
screen to achieve the desired result