Battery uptime
is low
No Live image
from the
visualization
device on the
displaying unit
screen
or
No light in the
visualization
device LED
16
Capacity of internal
battery has degraded
due to wear and tear.
Failure on visualization
device or visualization
device connection
The displaying unit
and the Ambu
visualization device
have communication
problems
Failure on visualization
device connector port
A recorded image is
shown in the
Procedure Folder tab
or
A menu box is blocking
the live image
If you experience that the battery life of the
displaying unit after a full charge is below 1
hour, consider replacing the battery.
Connect/reconnect visualization device
Restart the displaying unit
1.
Press the power button to turn OFF
the displaying unit
2.
When the displaying unit is OFF,
press the power button again
to turn it back ON
If the displaying unit does not turn OFF,
force it to turn OFF by pressing the power
button for 10 seconds. Press the power
button again turn the displaying unit
back on.
If this does not work reset displaying unit
hardware
1.
Remove all connected items including
the power supply
2.
Press the hardware reset button on
the back of the displaying unit*
3.
Reconnect the power supply
4.
Press the power button to turn it
back ON
Try the other visualization device
connector port.
If an image is obtained from the other
connector port, then the first connector
port may need to be replaced.
If none of the connector ports provide
an image, try again using a new
visualization device.
Return to live image by pressing
the Live View tab or
If this does not work restart
the displaying unit
1.
Press the power button to turn OFF
the displaying unit
2.
When the displaying unit is OFF,
press the power button again
to turn it back ON