procedure. Consult the tables below for guidance in positioning
the Visum Surgical Lights and StrykeCAM Overhead
Camera.
Guidance and Manufacturer's Declaration: Electromagnetic Emissions
Visum Surgical Lights and StrykeCAM Overhead Camera is intended for use in the electromagnetic
environment specified below. The customer or the user of Visum Surgical Lights and StrykeCAM Overhead
Camera should ensure that it is used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Caution Portable and mobile RF communications equipment may affect
the normal function of the Visum Surgical Lights and
StrykeCAM Overhead Camera.
Note
The Visum Surgical Lights and StrykeCAM Overhead Camera have
been designed and tested to comply with IEC 60601-1-2:2001
requirements for EMC with other devices.
Guidance and Manufacturer's Declaration: Electromagnetic Immunity
Visum Surgical Lights and StrykeCAM Overhead Camera are intended for use in the electromagnetic
environment specified below. The customer or the user of Visum Surgical Lights and StrykeCAM Overhead
Camera should ensure that it is used in such an environment.
IEC 60601 Test
Immunity Test
Level
Electrostatic
±6kV contact
Discharge (ESD)
±8kV air
IEC61000-4-2
Electrical fast
±2kV for power
transient/burst
supply lines
IEC61000-4-4
±1kV for input/
output lines
Power frequency (50/
3 A/m
60Hz) magnetic field
IEC 61000-4-8
NOTE: Ut is the a.c. mains voltage prior to application of the test level.
Compliance
Electromagnetic Environment - guidance
Group 1
Visum Surgical Lights and StrykeCAM Overhead Camera
uses RF energy only for its internal function; therefore, its
RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
Class B
Visum Surgical Lights and StrykeCAM Overhead Camera
are suitable for use in all establishments, including domestic
establishments and those directly connected to the public
low-voltage power supply network that supplies buildings
used for domestic purposes.
Compliance
Level
±2,4,6kV
contact
±2,4,8kV air
±2kV for
power supply
lines (Sidne)
±1kV for
input/output
lines
3 A/m
Electromagnetic Environment: Guidance
Floors should be wood, concrete, or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.
Mains power quality should be that of a
typical commercial or hospital environment.
Power-frequency magnetic fields should be
at levels characteristic of a typical location in
a typical commercial or hospital
environment.