Idiomas disponibles

  • MEXICANO, página 277
6. Pull on the end of the adjustable backrest strap to tighten.
7. Insert the mainframe straps between the product frame and the mattress.
8. Insert the buckle behind the litter cross brace and bring it up in front of the litter cross brace.
9. Fasten the buckle around the litter cross brace. Leave slack in the strap for final adjustment.
10. Tighten all straps.
N N o o t t e e - - For more information about the P P e e d d i i - - M M a a t t e e ® infant restraint system, see the manufacturer's instructions for use,
operation, and care. Safe and proper use of the P P e e d d i i - - M M a a t t e e ® infant restraint system is at the discretion of the user. Set
methods and procedures to educate and train your staff on the safe operation or maintenance of this product. These are
general instructions only. Keep these instructions for future reference. These instructions are a permanent part of the
product and should remain with the product even if the product is sold.
P P e e d d i i - - M M a a t t e e ® is a registered trademark of Ferno Washington, Inc.
A A t t t t a a c c h h i i n n g g t t h h e e d d e e f f i i b b r r i i l l l l a a t t o o r r p p l l a a t t f f o o r r m m
W W A A R R N N I I N N G G - - Do not attach the defibrillator platform on a Model 6086 P P e e r r f f o o r r m m a a n n c c e e - - P P R R O O XT cot with the P P o o w w e e r r - - L L O O A A D D
option to avoid the risk of a cot tip.
C C A A U U T T I I O O N N
• Always secure the defibrillator platform to the product when you use the defibrillator platform.
• Always use and adjust the straps that are provided with the defibrillator platform to secure the defibrillator.
• Always change the attachment location or adjust the straps for your specific defibrillator size or shape.
• Do not load the defibrillator platform above the safe working load of 30 lb (13.6 kg).
To attach the defibrillator platform:
1. Place the defibrillator platform in the stored position (Figure 24).
2. Open and expand the defibrillator platform legs (Figure 25).
F F i i g g u u r r e e 2 2 4 4 – – S S t t o o r r e e d d p p o o s s i i t t i i o o n n
F F i i g g u u r r e e 2 2 5 5 – – O O p p e e n n d d e e f f i i b b r r i i l l l l a a t t o o r r p p l l a a t t f f o o r r m m l l e e g g s s
3. If the product is equipped with an IV pole, raise the IV pole (A) to the up position (Figure 26).
4. Place the defibrillator platform on the product frame.
5. Position the inside legs (B) of the defibrillator platform toward the head end of the product (Figure 26).