Dräger Polytron Repeater WirelessHART 6700 IR WL Manual Del Usuario página 53

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 97
Byte
Bit
Description
16
0
OUT_OF_SERVICE
1
INSTALLATION_CALIBRATION
2
FAULTS_SENSOR
3
FAULTS_ELECTRONICS
4
M: Maintenance required
5
O: Out of specification
6
C: Function Check
7
F: Failure
17
4
PROGRAM_FLOW_ERROR
19
2
DEV_CONFIG_CHECK
Instructions for use
|
Dräger Polytron
Cause
Start sequence active.
Start sequence failed.
Adjustment or bump test active.
Sensor defective.
Incorrect power supply setting (battery or wired).
Serious device error.
Error in electronics.
Error in electronics.
®
6700 IR WL, Dräger Polytron
Remedy
Wait for start sequence.
A
Wait up to 15 minutes for the start sequence to
complete.
B
Restart the device.
C
Check the condition of the battery. Test with new
battery pack.
D
Contact DrägerService.
E
(if applicable) complete or repeat gas library
update.
Finish adjustment or bump test.
A
Restart the device.
B
Check the condition of the battery.
C
Contact DrägerService.
A
Check the setting of the power supply.
B
Restart the device.
C
Reset device to factory settings.
D
Contact DrägerService.
A
Restart the device.
B
Reset device to factory settings.
A
Reset device to factory settings.
B
Contact DrägerService.
®
®
Repeater WirelessHART
|
en
Troubleshooting
53