13.5
Battery pack
13.5.1 Battery charge
The gas warning device emits three battery charge warning thresholds during
operation.
Status
Measured
value
POWER_LOW
available
POWER_LOW,
available
POWER_CRITI-
CAL
POWER_LOW,
NaN
POWER_CRITI-
CAL
13.5.2 Changing the battery pack
For information on changing the battery pack, see "Set up battery operation",
page 45.
14
Settings
14.1
Sensor settings
Sensor settings are available in PolySoft.
Instructions for use
|
Dräger Polytron
LED display Time before
Explanation
complete
discharge
none
2-6 months
Order bat-
tery pack.
Yellow LED
Tempera-
Change bat-
flashes
ture depen-
tery pack.
dent approx.
1-3 weeks
(at least 24
operating
hours after
first occur-
rence).
Yellow LED
-
A shut-down
lights up
is possible at
continuously
any time.
®
6700 IR WL, Dräger Polytron
–
Setting the capture range
The capture range allows unwanted zero deviations to be suppressed. The
setting affects the output measured value.
The feature can be used to define the capture range and set the capture
range limits.
–
Gas settings
The feature can be used to set the target gas, LEL category, and measuring
unit.
14.2
Resetting the gas warning device to the
factory settings
This can be done no earlier than 10 minutes after turning on the gas warning
device.
It will reset device settings, data for integration in WirelessHART
sensor configuration, SIL passwords and Bluetooth
The feature is also available in PolySoft.
Procedure using the magnetic pen
Placement of the magnetic pen
0 s
Reset gas warning device to factory settings:
1. Hold the magnetic pen on the marking (figure A 9) for approximately
15 seconds until the blue LED flashes quickly.
2. Lower the magnetic pen for 7 seconds.
3. Put the magnetic pen back on the marking for 1 second (LED stops flashing)
and then take it off.
✓ The gas warning device will restart and is reset.
®
Repeater WirelessHART
|
Settings
®
network,
®
pairing PIN.
LED flashes
15 s
0 s
7 s
8 s
®
en