Only process chemicals that have been tested and
approved (e.g. VAH or FDA approval or CE mark) and
which are compatible with the product's materials
according to the chemical manufacturers' recommen-
dations may be used for reprocessing the product. All
the chemical manufacturer's application specifications
must be strictly observed. Failure to do so can result in
the following problems:
Visual material changes (such as fading or color
changes in titanium or aluminum). For aluminum,
the application/process solution only needs to be
pH >8 to cause visible surface changes.
Material damage (such as corrosion, cracks, breaks,
premature aging or swelling).
Do not use metal cleaning brushes or other abra-
sives that would damage the product surface and
could cause corrosion.
Further detailed advice on hygienically safe and
material-/value-preserving reprocessing can be
found at www.a-k-i.org, section Publications, Red
Brochure – Proper maintenance of instruments.
12.3 Preparation at the place of use
Remove all attached components from the product
(tool and accessories).
Remove the attachment from drill and reamer.
Remove any visible surgical residues as much as
possible with a damp, lint-free cloth.
Place the dry product in a sealed waste container
and forward it on for cleaning and disinfection
within 6 hours.
12.4 Preparation before cleaning
Before the first automatic cleaning/disinfection:
attach ECCOS fixations in a suitable sieve basket.
Insert the product in its proper position in the
ECCOS fixations.
Kirschner wire attachment GB643R
Turn adjustment sleeve 4.2 to setting for largest
Kirschner wire diameter.
16
12.5 Cleaning/disinfection
12.5.1 Product-specific safety instructions
for the reprocessing procedure
Damage to the product due to
inappropriate cleaning/disin-
fecting agents and/or exces-
CAUTION
sive temperatures!
Note
The indicated drying time is a guide time only. It must be
checked taking into account the specific conditions (e.g.
load) and if applicable adjusted.
Use cleaning and disinfect-
ing agents according to the
manufacturer's instruc-
tions which
be approved for plastic
material and high-
grade steel,
do not attack softeners
(e.g. in silicone).
Do not use cleaning agents
that contain acetone.
Observe specifications
regarding concentration,
temperature and exposure
time.
Do not exceed the maxi-
mum temperature of 60 °C
during chemical cleaning
and/or disinfection.
Do not exceed the maxi-
mum temperature of 96 °C
during thermal disinfection
with FD water.
Dry the product for at least
10 minutes at a maximum
of 120 °C.