14. Troubleshooting list
14.1 General malfunctions
Malfunction
Excessive noise
Tool not moving
Deficient cutting per-
formance
Tool cannot be cou-
pled
Detection
Excessive noise
Attachment can not be
extracted from the motor
unit
Motor drive not turning;
blocked
Motor running, but not
driving the saw
-
Tool overheats
Incompatible tool
Debris on tool or in cou-
pling
Deformed tools/coupling
components
Cause
Defective gearings
Attachment not cou-
pled properly
Defective gearings
-
Blunt tool
Wrong tool
Debris
Deformation
Remedy
Have product repaired by the
manufacturer
Couple the attachment cor-
rectly, see Chapter 11.1.2
Have product repaired by the
manufacturer
Have product repaired by the
manufacturer
See TA011869
Use a new tool
Insert appropriate tool
Clean the tool/coupling
Have product repaired by the
manufacturer