Descargar Imprimir esta página

B.Braun Aesculap GA861 Instrucciones De Manejo/Descripción Técnica página 15

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 52
7.8
Mechanical cleaning/disinfecting with manual pre-cleaning - tools
Note
The cleaning and disinfection device must be of tested and approved effec-
tiveness (e.g. FDA approval or CE mark according to DIN EN ISO 15883).
Note
The cleaning and disinfection device used for processing must be serviced
and checked at regular intervals.
Manual pre-cleaning with ultrasound and brush
Phase
Step
I
Immersion
II
Rinsing
III
Ultrasonic cleaning
IV
Cleaning with brush
D–W:
Drinking water
RT:
Room temperature
*Recommended: BBraun Helizyme
**Recommended: BBraun Stabimed
Only perform ultrasound cleaning in the ECCOS holder GB718R to pre-
vent any damage to the cutting edges.
Phase I
Fully immerse the product in enzymatic cleaning solution for at least
30 minutes. Ensure that all accessible surfaces are moistened.
Phase II
Rinse the entire product (all accessible surfaces) under running water.
Phase III
Tilt up the tool holder 10 of the ECCOS holder GB718R by the side
handles 8, see Fig. C.
Clean the product in an ultrasonic cleaning bath (frequency 35 kHz) for
at least 15 minutes. Ensure that all accessible surfaces are immersed
and acoustic shadows are avoided.
Phase IV
Clean the product with a suitable cleaning brush until all discernible
residues have been removed.
T
t
Conc.
[°C/°F]
[min]
[%]
RT (cold) >30
50
RT (cold) -
-
55/131
>15
2
RT (cold) -
-
Water
Chemicals
quality
D–W
Enzymatic detergent*
D–W
-
D–W
Concentrate-free, aldehyde-free, phenol-free, QUAT-free, pH ~ 9**
D–W
-
13

Publicidad

loading

Este manual también es adecuado para:

Aesculap ga864Aesculap ga865Aesculap ga869