6c2. Alarm list menu
In View measure mode, press
access the main menu, use
Alarm list and press
6c3. Calibration menu
In View measure mode, press
access the main menu, use
Calibration and press
The Calibration menu consists of fi ve (5) sub-menus:
A. pH probe
B. ORP probe
C. Temperature probe
D. Flow rate sensor
E. Free Chlorine probe
Use
to select an item and press
A. pH Probe Calibration menu
The pH Probe Calibration menu consists of four (4) sub-menus:
• Automatic: the instrument requires the standard buffer solutions 7 pH, 4
pH or 9.22 pH.
• Manual: the instrument will suggest the buffer solutions from the default
values, but the value can be changed.
• Reference: the instrument accepts the calibration of one point with a
manually set value.
• Reset (Calibration): the calibrations can be deleted and restored the
default values.
Scroll through the menu using
for 3 seconds to
to select the item
to confi rm.
for 3 seconds to
to select the item
to confi rm.
to confi rm.
, select the item and confirm with
ProChem Double | pH · ORP · Chlorine
Only use spare parts supplied by Hayward brands ®
Main Menu
□ Calibration
□ Setup
□ Advanced
■ Alarm list
□ Info/Priming
Main Menu
■ Calibration
□ Setup
□ Advanced
□ Alarm list
□ Info/Priming
.
Page 17 on 42
Alarm list
■ Log items
■ Reset Alr Relay
■ Reset OFA
Calibration
■ pH
□ mV
□ Temp
□ Flow Rate sensor
□ Chlorine
pH Probe
■ 2 point
□ 1 point
□ Reference
□ Cal. Reset