2. Remove eventual air bubble in the membrane cap by
gently tapping the cap using the body of the sensor
3. Clean the electrode using the provided abrasive paper.
Note: the electrode must be a little inclined on the abra-
sive paper and not perfectly vertical, in order not ot
remove the brown film that is on the extremity.
4. Assemble the sensor by inserting the body in the mem-
brane cap and screwing it clockwise until it is correctly
insertd. Exxcesive electrolite will overflow from the valve
positioned where the red arrow is, you can clean by
using simple water.
5. Move the second silicone gasket on top of the first one.
6. Insert the terminal cap onto the cable through the PG7
connector and fix the cable wir into the terminals. The
recommended cable is: 4mm diameter, 2x 0.25 mm²
wires.
ProChem Double | pH · ORP · Chlorine
Only use spare parts supplied by Hayward brands ®
Page 8 on 42