Parameter
Range of values
2nd Shock for VT1/VT2
OFF; 4 ... (2) ... 20 ... (5) ... 40 J 40 J
3rd - nth shock for VT1/VT2
OFF; 4*40 J; 6*40 J
1st Shock for VF
2 ... (2) ... 20 ... (5) ... 40 J
2nd Shock for VF
4 ... (2) ... 20 ... (5) ... 40 J
3rd - nth Shock for VF
4*40 J; 6*40 J
For shock in VT1/VT2 and VF:
– Confirmation
OFF; ON
– Polarity
Normal; Reverse; Alternating Normal
– Shock form
Biphasic; Biphasic 2
– Shock path
RV -> ICD+SVC
RV -> ICD
RV -> SVC
Sensing
Sensitivity and thresholds
Parameter
Range of values
Sensing A
STD; OFF
Sensing RV
STD; TWS; VFS; IND
Sensing LV
STD; OFF; IND
Upper threshold RV
50; 75%
Upper threshold LV
50; 75%
Upper threshold duration
110; 150 ... (50) ... 500 ms
after detection
VFS: 110 ms
Upper threshold duration
after pacing
Lower threshold RV
25; 50%
T-wave suppression
OFF; ON
after pacing
Minimum threshold A
0.2 ... (0.1) ... 2.0 mV
Standard
Parameter
x x x x
Minimum threshold RV
6*40 J
x x x x
Minimum threshold LV
40 J
x x x x
40 J
x x x x
Diagnostics
6*40 J
x x x x
The following can be set:
Parameter
ON
x x x x
For AT/AF
Biphasic
For SVT
RV->
x
x x
Periodic recording
ICD+SVC
RV -> ICD
x
IEGM configuration
The following can additionally be set for Ilesto 7:
Standard
Parameter
Std.
x x x
Start resting period
Std.
x x x x
Std.
x
Duration of resting period
50%
x x x x
AV delay modification in
50%
x
sensing test
350 ms
x x x x
Thoracic impedance (TI)
400 ms
25%
x x x x
OFF
x x x x
0.4 mV
x x x
18
Range of values
Standard
0.5 ... (0.1) ... 2.5 mV
0.8 mV
0.5 ... (0.1) ... 2.5 ... (0.5) ... 5.0
1.6 mV
mV
Range of values
OFF; ON
x x x ON
for Ilesto 7: Extended ON
OFF; ON
x x x ON
When Home Monitoring OFF:
x x x x 90 days
OFF; 30 ... (30) ... 180 days
RA, RV, LV
RA, RV, FF
FF; RV; LV
Range of values
0:00 ... (1:00) ... 23:00 hh:mm x x x x 2:00
0.5 ... (0.5) ... 12 h
x x x x 4 h
OFF; 300 ms
x x x 300 ms
OFF, ON
x x x x OFF
x x x x
x
Standard
x RA, RV,
LV
Standard
hh:mm