Telemetry data for Japan:
In accordance with Japanese law, this device has been assigned an identification
number under the "Ordinance concerning certification of conformity with technical
regulations etc. of specified radio equipment", Article 2-1-8.
• R: 202-SMA026
Pulse form
The pacing pulse has the following form:
The pulse amplitude reaches its maximum value at
the beginning of the pulse (Ua). With increasing
pacing duration (tb), the pulse amplitude is reduced
dependent on the pacing impedance.
Resistance to interference
• Note on device type DX (only devices with a DF-1/IS-1 connection): The EMC
requirements are met as long as atrial sensitivity is set to 1.0 mV (factory settings)
or values ≥ 1.0 mV. Measures must be taken to assure interference-free therapy if
more sensitive values are set.
• Note on device type HF: In the case of unipolar sensing, the requirement for inter-
ference voltages of ≤ 0.3 mV (peak to peak) is met.
Common mode rejection ratio
Rate
Common mode rejection ratio
Atrium: DX*
Atrium: DR, HF
16.6 Hz
58 dB
53 dB
50 Hz
55 dB
55 dB
60 Hz
56 dB
56 dB
* only devices with a DF-1/IS-1 connection.
ATP amplitude
A burst was measured at 500 Ω, an amplitude of 7.5 V (tolerance ±1.5 V), pulse width of
1.5 ms, R-S1 interval of 300 ms and an S1 count of 5:
ATP amplitude
Measured
minimum
RV
7.67 V
LV
7.67 V
V right: VR, DR, HF
V left: HF
64 dB
66 dB
64 dB
66 dB
64 dB
68 dB
Measured
Mean value
maximum
7.67 V
5.00 V
7.67 V
4.99 V
Automatic sensitivity setting
Measurement of actual values and test signal wave shape: standard triangle. For the
device type DX, the programmed atrial sensitivity is intensified by a factor of 4.
Sensitivity
Value
A: positive
0.2 mV
A: Negative
DX: A: positive
0.2 mV
DX: A: negative
RV: positive
0.5 mV
RV: negative
LV: positive
0.5 mV
LV: negative
Shock energy / peak voltage
With shock path: RV to housing + SVC
Shock energy
Tolerance peak
(Tolerance)
voltage
1 J (0.7 ... 1.18)
90 ... 120 V
20 J (16.9 ... 20.9)
440 ... 480 V
40 J (33.8 ... 41.4)
620 ... 690 V
Battery Data
Battery characteristics
The following data is provided by the manufacturers:
Manufacturer
Battery type
Battery ID number shown on the
programmer
Device type
Battery voltage at ERI
20
Tolerance
Measured value
0.2 ... 0.5
0.24 mV
0.24 mV
0.2 ... 0.52
0.05 mV
(0.05 to 0.13)
0.05 mV
0.3 ... 0.7
0.48 mV
0.40 mV
0.3 ... 0.7
0.48 mV
0.56 mV
Measured value
Measured value
Shock energy
Peak voltage
0.84 J
100 V
18.1 J
469 V
36.9 J
667 V
GREATBATCH, INC.
LITRONIK GmbH & Co
Clarence, NY 14031
01796 Pirna, Germany
GB 2992
LiS 3410 RA/RR
3
4
VR, (DX), DR, HF
2.5 V
2.85 V