Fill Level Indicator and Remaining Amount
531
Fill Level Indicator
The fill level indicator (1) helps you to estimate the amount of impression material remaining
in the cartridge. However, after installing a new cartridge or if the cartridge is very full, the red indica-
tor will not be visible until the cartridge has started to be used.
532
Remaining Amount Detection
The Dynamix speed is equipped with an electronic fill level detection. Once there is
only sufficient impression material remaining in a cartridge to fill approx. one casting spoon,
the yellow "Residual level" LED (5) goes on.
The indicator passes a residual level marking.
GB
6
Cleaning / Maintenance
- 8 -
To clean the unit, wipe it down with a moist cloth.
Never use cleansers containing solvents.
61
Cleaning the Interior
– Cartridge bed (14, fig. 1): With the plungers (9, fig. 1) retracted, the cartridge bed is easily accessible for cleaning.
– Plungers' disc (9, fig. 1) and spindle covers (10, fig. 1): With the unit cover closed and without a cartridge inserted the
pistons may be moved into a position suitable for removing residues on the plungers' disc or the spindle covers.
NEVER clean the spindles (11, fig. 1) themselves. The grease applied to spindles is necessary for the
unit's proper function.
Tip: Remnants of impression material are most easily removed with a dry cloth or paper towel.
62
Disinfection
Never use superheated steam for disinfection!
Use only appropriate disinfectants for disinfection. Please pay attention to the manufacturer's instructions
for the disinfectant used. If necessary, dry housing parts or exposed connectors after disinfecting.
63
Plungers' Disc Replacement
The plungers' disc (9, fig. 1) are subject to wear due to rubbing against the inner wall of the cartridge (20, fig. 1).
Plungers' disc that are much worn may damage the foil bag in the supporting cartridge (20, fig. 1).
Replacement plungers' disc are therefore available as a spare part.
The tool required for the replacement and a description of the procedure are enclosed with the spare part.
64
Unit Cover Replacement
The unit cover (8, fig. 1) is available as a spare part.
The tool required for the replacement and a description of the procedure are enclosed with the spare part.
65
Fuse Replacement
Before replacing the fuse, disconnect the unit from the main power supply by unplugging the power cord
from the wall outlet.
Only replace fuses with new ones that have the same rating as those being replaced.
Refer to the "Technical Specifications" section.
Never use fuses with a higher rating.
• Unplug the unit.
• Pull the fuse holder (17) out.
• Remove the defective fuse and insert a new one.
• Push the fuse holder back in, making sure it is completely seated.