Replacing Fuses; Treatment; Disassembly; Execution Instructions - Krömker UltraNeb Instrucciones De Uso

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

8.3

Replacing fuses

D WARNING!
Electric shock!
Disconnect the device from the mains before
replacing the fuses.
3
2
4
1
Device socket
2
Fuses
3
Locking mechanisms
4
Fuse insert
D WARNING!
Observe fuse type!
Only fuses of the following type may be used:
T1.0 A H 250V 5x20 mm
To replace the fuses in the device socket, proceed as
follows:
1.
Unplug the mains cable from the device
socket (1).
2.
Carefully unlock the side locking mechanisms (3)
on the device socket (1) with a small screwdriver
and remove the fuse insert (4).
3.
Remove defective fuses (2), replace with new
fuses.
4.
Insert fuse insert (4) and press into place until it
audibly latches.
5.
Check secure seating of the fuse insert (4).
48
9.
9.1
☞ Unplug mains plug and device plug.
☞ Unplug the tube heating plug.
☞ Remove tubes and bacteria filter.
☞ Slide the retainer up and remove the cover from
☞ Pull the nebuliser chamber out of the connection
☞ Dispose of disposable items (disposable tubes and
☞ Do not disinfect or sterilise filters.
☞ Replace and dispose of filters according to the
9.2
The individual parts of the ultrasonic nebuliser
can be cleaned and disinfected according to the
specifications in the following table. When doing so,
1
please observe the following points:
− Before disinfection the parts must always be
− Dirt particles can become encapsulated and
− When using disinfectants, please follow the
Danger of injury!
Disinfectant can contain substances that are
hazardous to health and that can cause irritation
upon contact with skin or eyes. Protect your skin and
eyes and observe the hygiene rules when working
with disinfectants.

Treatment

Disassembly

the nebuliser chamber.
and pour out liquid from the nebuliser chamber.
Observe the hygiene regulations of the hospital!
covers, containers of sterile water systems) in an
environmentally friendly manner. Observe hygiene
regulations.
manufacturer's instructions.

Execution instructions

cleaned and then dried.
cause equipment parts to be non-sterile after
disinfection. Therefore, all parts of the device
must be thoroughly cleaned before disinfection
and stubborn dirt must be removed with a cloth
and detergent.
instructions of the manufacturer and the hygiene
specialist.
D WARNING!
96200-114 – 10 – 15.07.2020
Treatment