Cleaning in a washing machine
Rinse cycle up to 95° C:
− Nebuliser chamber (transparent)
− Nebuliser chamber cover (transparent)
Not permissible:
− Housing
− Mains cable
− Accessory kits
Autoclaving
Attention, hot steam accelerates the natural ageing
of plastics. The function of the device parts can be
impaired by the change in material.
Up to 121° C, max. 20 min
− Nebuliser chamber (transparent)
− Nebuliser chamber cover (transparent)
Not permissible:
− Housing
− Mains cable
− Accessory kits
Examples of surface disinfectants:
− Incidin® Plus *
− Incidin® Perfect *
− Antiseptica Kombi surface disinfectant *
− Antifect® FF *
− B 10 *
Examples of instrument disinfectants:
− Cidex® AF
− Gigasept® FF *
* registered trademarks of the respective
manufacturers
The lifespan of the nebulizer chamber with lid and
the quartz module is limited to 30 autoclaving cycles.
The aforementioned example disinfectants are
recommendations from the device manufacturer. The
recommendation of the mentioned active ingredient
bases (pH 4.5-8) applies. This does not release the
user from the need to consult a hygiene specialist
in individual cases or to test the components to
be disinfected for compatibility with the active
ingredients.
50
9.4

Manual cleaning of the quartz

☞ To clean, remove the nebuliser chamber cover (7)
from the nebuliser chamber (6) and unscrew the
quartz module (9) from the nebuliser chamber (6).
Reassembly takes place in reverse order.
9.5

Cleaning run-on kits

☞ Please refer to the separate user manual for
instructions.