Treatment
D ATTENTION!
Material changes!
Almost all components of the ultrasonic nebuliser
are made of plastics. Solvents, some disinfectants
or cleaning agents can dissolve the plastics or cause
stress cracks. Therefore do not use any agents
containing hydrocarbons or alcohol to clean the
surfaces.
D ATTENTION!
Function testing!
After cleaning or disinfection, always check that all
parts of the device are in good working order.
D „7.1 Function test" (Seite 45).
NOTE:
Discoloration of the plastic parts may occur when
using non-colourfast wrapping cloths.
NOTE:
After cleaning, disinfection and complete drying,
always check the rubber seal of the quartz module
carefully for possible damage.
96200-114 – 10 – 15.07.2020
9.3
Performing cleaning
General
Rinse the nebuliser chamber and nebuliser chamber
cover with distilled water after each use. Disinfection
must be carried out with standard medical
disinfectants.
The quartz module remains firmly screwed into the
nebuliser chamber.
Rinse all individual parts with distilled water after
each use.
Disposable item
Intended for single use only.
Disposable items must not be re-treated, there is a
risk of infection.
− Bacteria and virus filter:
Replace after 1 week at the latest or according
to special instructions from the manufacturer.
When using a closed sterile water system, it is
recommended to change the filter after 48 hours
of operation or every 14 days.
Cleaning
In warm water with mild household cleaner:
− Nebuliser chamber (transparent)
− Nebuliser cover (transparent)
Wipe with a damp cloth:
− Housing
− Mains cable
− Accessory kits
Disinfection
After the prescribed exposure time (according to the
manufacturer's instructions), components must be
cleaned of disinfectant residues with a damp cloth
and then dried.
Immerse in diluted solution:
− Nebuliser chamber (transparent)
− Nebuliser chamber cover (transparent)
Wipe disinfection:
− Housing
− Mains cable
− Accessory kits
49