Hurrimix
is designed to be used in the electromagnetic environment specified below. The client or user must ensure that it is used in such an environment.
2
IMMUNITY
IEC 60601-2 level
test
Conducted
3 Vrms
RF
150 kHz to 80 MHz
IEC 61000-4-6
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2.5 GHz
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and
people.
a) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM
radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters,
an electromagnetic site survey should be considered. If the measured field strength in the location in which Hurrimix
compliance level above, the product's normal operation should be observed. Additional measures, such as reorienting or relocating the product, may be
required.
b) The field strength over a frequency ranging from 150 kHz to 80 MHz should be less than 3 V/m.
Hurrimix
is designed to be used in the electromagnetic environment specified below. The client or user must ensure that it is used in such an environment.
2
IMMUNITY
EN 60601-2 level
test
Conducted
3 Vrms
RF
150 kHz to 80 MHz
EN 61000-4-6
Radiated RF
3 V/m
EN 61000-4-3
80 MHz to 2.5 GHz
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and
people.
a) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM
radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters,
an electromagnetic site survey should be considered. If the measured field strength in the location in which Hurrimix
compliance level above, the product's normal operation should be observed. Additional measures, such as reorienting or relocating the product, may be
required.
b) The field strength over a frequency ranging from 150 kHz to 80 MHz should be less than 3 V/m.
hurrimix
2
hurrimix
2
115Vac/60hz
Manufacturer's declaration and guidance – electromagnetic immunity
Compliance
level
Portable and mobile RF communications equipment should be used no closer to any of Hurrimix
parts, including cables, than the recommended separation distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance
d = [3,5 / V
] √P
1
d = [3,5 / E
] √P from 80 MHz to 800MHz
1
V
= 3 V rms
d = [7/ E
]
√P from 800 MHz to 2,5 GHz
1
1
where P is the maximum output power rating of the transmitter in Watts (W), according to the
transmitter manufacturer, and d is the recommended separation distance in meters (m).
E
= 3 V / m
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey
1
should be less than the compliance level in each frequency range
Interference may occur in the vicinity of equipment marked with the following symbol:
hurrimix
230Vac/50hz
2
Manufacturer's declaration and guidance – electromagnetic immunity
Compliance
level
Portable and mobile RF communications equipment should be used no closer to any of Hurrimix
parts, including cables, than the recommended separation distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance
d = [3,5 / V
] √P
1
d = [3,5 / E
] √P from 80 MHz to 800MHz
1
V
= 3 V rms
d = [7/ E
]
√P from 800 MHz to 2,5 GHz
1
1
where P is the maximum output power rating of the transmitter in Watts (W), according to the
transmitter manufacturer, and d is the recommended separation distance in meters (m).
E
= 3 V / m
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey
1
should be less than the compliance level in each frequency range
Interference may occur in the vicinity of equipment marked with the following symbol:
The reference figures are present on pages 3, 4, 5 and 6 of this manual and in the product Quick Guide.
Electromagnetic Environment - Guidance
2
is used exceeds the applicable RF
Electromagnetic Environment - Guidance
is used exceeds the applicable RF
2
's
2
(a)
,
.
(b)
's
2
(a)
,
.
(b)
25
EN