Riester ri-medic Manual De Instrucciones página 16

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Press the Manual mode manual button to re-inflate the cuff.
Press the Start/Stop button if you need to rapidly release all the air from the cuff.
-It is recommended to use the phase V of the Korotkoff- sounds (K5) for auscultatory measurement at adults.
-It is recommended to use the phase IV of the Korotkoff- sounds (K4) for auscultatory measurement at children
age 3 to 12 years.
-It is recommended to use the phase V of the Korotkoff- sounds (K5) for auscultatory measurement at pregnant
female patient, except that Korotkoff- sounds are heard when cuff deflated , in this case, K4 should be used.
5.4 MANAGING READINGS
a.) Recalling the Last Set of Readings
To redisplay the last set of readings, depress the Start/Stop button on the BP module for more than two seconds until the last
reading is displayed. If the last attempted reading resulted in an error and/or warning, then this will be displayed. The device
will display dashes if no reading is in memory, a reading was aborted or, the previous BP was a Manual mode measurement.
b.) Clearing the Last Set of Readings
To clear the values from the last automatic BP measurement and the accessories, press and hold the Start/Stop button more
than 5 seconds. Previous values will be displayed momentarily. Then the display blanks. On redisplay you will see dashes for
all the values that have been cleared.
Your Riester ri-medic displays the most recent set of readings for two minutes. If patient privacy is a concern,
you can clear these readings from the display before collecting vital signs from another patient.
5.5 BLUETOOTH® WIRELESS TRANSMISSION OF READINGS
If your Riester ri-medic has the optional Bluetooth® wireless communication capability you may be able to transmit the rea-
dings into your EMR/EHR system or communication network. Contact your IT administrator or EMR/EHR vendor for support
on configuring your Riester ri-medic for Bluetooth® wireless communication.
6. MAINTAINING THE RIESTER ri-medic
a.) Routine Maintenance
Establishing simple care guidelines helps protect the performance and life of your Riester ri-medic. On a routine basis, you
should inspect the device, cables and pneumatic hoses for cracks, fraying or kinks and immediately replace any damaged
parts.
Remember to check the calibration of the BP module every two years. If available, a biomedical technician may help in
maintaining your equipment.
b.) Cleaning
Cleaning the Device
Wipe the device with a soft, damp cloth to remove surface dust and dirt.
The Riester ri-medic device cannot be sterilized.
Never immerse the device in any fluid or attempt to employ cleaning fluids or solvents.
c.) Cleaning the Cuffs
Between uses, wipe cuff sleeves and the insides of cuffs with a medical grade cleaning agent.
Periodically, remove the bladders and machine-wash the cuffs in cold water.
Line dry.
16

Publicidad

Tabla de contenido
loading

Tabla de contenido