Attention!
• Never place the instrument heads and handles in liquids! Make sure that no
liquids penetrate the housing interior!
• The article is not approved for machine reprocessing and sterilisation. This
can lead to irreparable damage!
17.3. Sterilisation
a) Reusable ear specula
The ear specula can be sterilised at 134° C, with a 10 minute period in the
steam steriliser.
b) Single use ear specula
For single use only
Attention: Repeated use lead to infection.

18. Spare parts and accessories

A detailed list can be found in our brochure, „Instruments for H.N.O.," Ophthal-
mologic Instruments, which you can find at www.Riester.de https://www.Ries-
ter.de/en/productdetails/d/ri-scoper-l-premium-ent-and-ophthalmic-inst-
ruments/ri-scoper-l-otoscopes/

19. Maintenance

The instruments and their accessories require no special maintenance. If an
instrument needs to be inspected for any reason, please send it to us or to an
authorised Riester dealer in your area, whom we will gladly name upon request.

20. Instructions

Ambient temperature:
Relative humidity:
Transport and storage temperature: -10 ° to +55 °
Relative humidity:
Air pressure:

21. ELECTROMAGNETIC COMPATIBILITY

ACCOMPANYING DOCUMENTS ACCORDING TO
IEC 60601-1-2, 2014, Ed. 4.0
Attention:
Medical electrical equipment is subject to special precautions regarding elec-
tromagnetic compatibility (EMC).
Portable and mobile radio frequency communication devices can affect medical
electrical equipment. The ME device is for operation in an electromagnetic en-
vironment or home health care and intended for professional facilities such as
industrial areas and hospitals.
The user of the device should ensure that it is operated within such an en-
vironment.
Warning:
The ME device may not be stacked, arranged or used directly next to or with
other devices. When operation is required to be close to or stacked with other
devices, the ME device and the other ME devices must be observed in order to
ensure proper operation within this arrangement. This ME device is intended
for use by medical professionals only. This device may cause radio interference
or interfere with the operation of nearby devices. It may become necessary to
take appropriate corrective measures, such as redirecting or rearranging the
ME device or shield.
The rated ME device does not exhibit any basic performance features in the
sense of EN60601-1, which would present an unacceptable risk to patients,
operators or third parties should the power supply fail or malfunction.
Warning:
Portable RF communications equipment (radios) including accessories, such as
antenna cables and external antennas, should not be used in closer proximity
than 30 cm (12 inches) to parts and cables of the ri-scope L instrument head
specified by the manufacturer. Failure to comply may result in a reduction in the
device's performance features.
Directives and manufacturer's declaration - Electromagnetic emissions
The ri-scope L instruments are intended for use in the electromagnetic environ-
ment specified below. The customer or user of the ri-scope® L should ensure
that it is used in such an environment.
26
0° to +40°
30 % to 70 % non-condensing
10 % to 95 % non-condensing
800 hPa - 1100 hPa