Guidance and manufacturer's declaration - electromagnetic emission
ri-‐scope
L
The
i nstrument is intended for use in the electromagnetic environment specified below. The
customer or the user of the e-scope® should ensure that it is used in such an environment
Emission test
RF emissions
HF- emissions pursuant to
CISPR 11
RF emissions
HF- emissions pursuant to
CISPR 11
Emissions of harmonics
IEC 61000-3-2
Emissions of voltage
fluctuations, flicker
IEC 61000-3-3
Guidance and manufacture's declaration – electromagnetic immunity
ri-‐scope
L
The
i nstrument is intended for use in the electromagnetic environment specified below. The customer or the
ri-‐scope
L
user of the
s hould ensure that it is used in such an environment
Immunity testing
Electrostatic discharge
Con: ±8 kV
(ESD)
Air: ±2,4,8,15 kV
IEC 61000-4-2
Fast transient electrical
5/50 ns, 100 kHz , ±2 kV
disturbances / bursts IEC
61000-4-4
Surge voltage
± 0.5 kV voltage
Phase-to-phase conductor
± 2 kV voltage
IEC 61000-4-5
Line-to-earth
Voltage dips, short-term
<0% UT 0.5 period at 0.45, 90, 135,
interruptions and
180, 225, 270 and 315 degrees
fluctuations in the supply
voltage acc. to
0% UT 1 period and 70% UT 25/30
periods single-phase: at 0 degrees
IEC 61000-4-11
(50/60 Hz)
Magnetic field with
30A/m
efficiency-rated
frequencies
50/60 Hz
IEC 61000-4-8
NOTE UT is the AC source. Mains voltage before the application of the test level.
Compliance
The
exclusively for an internal function. Therefore
Group 1
his RF transmission is very low and it is
unlikely to be adjacent
electronic devices are disturbed.
The
Class B
establishments, including residential areas and
those directly connected to a public supply network
that also supplies buildings used for residential
purposes.
Not applicable
Not applicable
IEC 60601 test level
Not applicable
Not applicable
Not applicable
30A/m
50/60 Hz
Electromagnetic environment – guidance
ri-‐scope
L
u ses RF energy
ri-‐scope
L
i s intended for use in all
Compliance
Electromagnetic environment -
Instructions
Floors should be wood, concrete or
Con: ±8 kV
ceramic tile. If floor are covered with
Air: ±2,4,8,15 kV
synthetic material, the relative
humidity should be at least 30%.
The quality of the supply voltage
should be that of a typical business
or hospital environment.
The quality of the supply voltage
should be that of a typical business
or hospital environment.
The quality of the supply voltage
should be that of a typical business
or hospital environment.
Mains frequency magnetic fields
should be at a level characteristic of
a typical location in a typical
commercial hospital environment.