4. Fully insert the hex wrench (packaged with the neurostimulator) into the self-sealing
grommet of the connector block and tighten each setscrew.
#
Cautions:
To avoid undertightening, do not use a torque wrench. Undertightening the
neurostimulator setscrews may result in insufficient electrical contact within the
connector block, resulting in intermittent or loss of stimulation.
Be sure the hex wrench is fully inserted into the self-sealing grommet. If the hex
wrench is not fully inserted, the setscrew may be damaged, resulting in
intermittent or loss of stimulation.
Before tightening setscrews, ensure that the extension connector pins are
inserted into the connector block to prevent damaging the connector block.
When using the hex wrench, do not overtighten the setscrews. Overtightening
may damage the setscrews and prevent setscrew removal.
Verify that each leaf of the self-sealing grommet is closed after the hex wrench
is withdrawn. If fluid leaks through a grommet seal that is not fully closed, the
patient may experience shocking, burning, or irritation at the neurostimulator
implant location, or intermittent stimulation or loss of stimulation may occur.
Note: The hex wrench must be oriented to the same angle as the setscrew (Figure 2).
Figure 2. Tightening the setscrews in the self-sealing grommet.