Table 1. Shipping, operating, and power-on-reset (POR) values for the
Model 7425 Itrel 3 Neurostimulator
Programmable parameter
Amplitude
Normal resolution
Upper limit
Lower limit
Fine resolution
Upper limit
Lower limit
Rate
Upper limit
Lower limit
Pulse width
Upper limit
Lower limit
Operation mode
Cycle ON/OFF Time
W/O SoftStart/Stop
With SoftStart/Stop
SoftStart/Stop
Ramp
Dose time
Dose lockout time
c
Output On/Off
Magnet control
6 English
7425 2004-05
221656002
Rev A
Medtronic Confidential
a
.
Shipping
Operating
0.0 V
0.1 V steps
0.0 V
10.5 V maximum
0.0 V
0.0 V minimum
-----
0.05 V steps
-----
6.35 V maximum
-----
0.0 V minimum
31 Hz
49 values
(from 2.1 to 130 Hz)
31 Hz
130 Hz maximum
2.1 Hz
2.1 Hz minimum
210 µs
increments of 30 µs
450 µs maximum
210 µs
60 µs
60 µs minimum
Continuous
Continuous or cycling
0.1 s
0.1 s to 24 h
-----
1 s to 24 h
Off
1 s, 2 s, 4 s, 8 s, or off
Off
15, 20, 25, 30 s, or off
at start of stimulation
Off
15, 30, 45, 60, 75 m,
or off stimulation
periods
0 h
1.0, 1.5, 2.0, 2.5, 3.0,
3.5, or 4.0 h
Off
On or off
On
On or off
NeuroImplant_R01
b
POR
0.0 V
0.0 V
0.0 V
-----
-----
-----
31 Hz
31 Hz
2.1 Hz
210 µs
210 µs
60 µs
Continuous
0.1 s
-----
Off
Off
Off
0 h
Off
On