of the sterilizer chamber. Steam can lift soil and poorly rinsed chamber
cleaning detergents from the chamber walls and deposit them onto
instruments and wrappers.
Rust Deposits: It is very unlikely for surgical grade steel to rust. What
appears to be rust is often residual organic matter in box locks or mineral
deposits which have been baked onto the surface of the instrument. In
localities where the water has a high iron content, for example, an iron
deposit will result in a metallic film on the instrument. This may be
prevented with the use of "treated water" for the FINAL rinse during
cleaning procedures.
The most effective method of dealing with instrument problems is to
prevent them from occurring. The use of "treated water", careful
preliminary cleaning, using neutralized pH solutions, following
manufacturer's instructions, and visual inspection, will help to keep
instruments performing accurately and cosmetically free of troublesome
stains. It is important to act quickly should a problem arise. Delay will
compound the problem and irreparable harm may result.