The setting is valid for continuous transmission in calibration unit and in current unit activated by
means of the balance (see parameter P4.2.1.2), it is also valid for continuous transmission activated
by command sent from a computer (see point 15 COMMUNICATION PROTOCOL).

13.1.4. Cooperation with E2R

E2R is a control system, it controls all weighing processes performed on a balance. Upon activation
of this option, operations on some Databases are possible only with the use of a computer (options
are not accessible on a balance). In order to start cooperation with E2R system set <YES> value of
<E2R SYSTEM> parameter.
CAUTION!
<E2R System> parameter activation may be carried out by the device manufacturer or authorized persons,
exclusively. Rememeber to switch continuous transmission on to provide correct coperation with the E2R
system.

13.1.5. Printout

Type of the printout sent to the port chosen for the computer.
Procedure:
− press
button
− enter <P4 DEVICES> menu
− enter menu group <P4.1 COMPUTER>
− enter <P4.1.7 PRINTOUT> parameter
− select printout:
NONE – no printout selected
NSD. PRN. 1÷4 – one of non-standard printouts (see: non-standard printouts
description)
MOB. APP – special printout (frame with data sent from the balance) recognized by
application used for previewing measurement result on mobile devices.

13.2. PRINTER

Submenu <PRINTER> allows the user to choose a port and a device, to which the data is to be sent
upon pressing
<PRINTOUTS/PRINTOUT GLP>.
Procedure:
press
button
-
enter <P4 DEVICES> menu
-
enter menu group: <P4.2 PRINTER>
-
select port to which the printer is to be connected:
-
COM 1 or COM 2 – RS 232 port, to which the printer is connected
USB 1 type A – USB port, to which PCL printer or EPSON printer is connected
WIFI – WIFI port, which can send data to special software manufactured by RADWAG
e.g. WIN measurement, run by means of the computer connected to the balance via
WIFI.
PENDRIVE – USB port type A, to which external memory flash drive is connected
USB PC –USB port type B, to which the computer with the special software by RADWAG
is connected e.g. WIN measurement.
USB FREE LINK – USB type B port, to which the computer is connected. The tool, acting
like a keyboard, is used for entering data for peripherals. On proper modification of the
printout and pressing declared balance button (e.g. F1 header, F3 footer, ENTER
printout), the data is directly entered to computer programs, e.g. Excel, Word, Notepad
etc.
button on a balance. Content of data being transferred is set in parameter