Transferred Data Format; Format Of Data Sent On Pressing Print Button; Response Format - RADWAG R Serie Manual Del Usuario

Ocultar thumbs Ver también para R Serie:
Tabla de contenido

Publicidad

Balance transmission parameters must correspond to the parameters of a device that is
connected to the balance.
Baud rate
Parity control
Value indicated on a display can be sent via RS232 or USB port to a peripheral device in one of four
accessible ways:
manually
- on pressing
automatically - on stabilisation of a weighing result
continuously - on activation of a function or sending a command
on command sent from a peripheral device (see additional functions).
Value indicated on a display can be sent via COM port or USB port in the following form:
stable
– data is sent immediately on stabilisation of weighing result (button
unstable – on pressing
a printout such status is marked with <?> symbol located in front of the weighing result). This
option is only available for non-verified balances.

14.1. TRANSFERRED DATA FORMAT

The measurement result can be transferred from a balance to an external device by pressing
button, located on the balance, or by sending a control command from a computer.

14.2. FORMAT OF DATA SENT ON PRESSING PRINT BUTTON

CAUTION!
Unstable measurement printout is disabled for a verified balance.
Printout format:
1
2
Stability marker
Character
Mass
Unit

14.2.1. Response Format

On receipt of a command, the balance responds as follows.
first:
XX_A CR LF
- 4800 ÷ 115200 bit / s
- NONE, ODD, EVEN
button
button, display status is sent immediately to a peripheral device (on
3
4 - 12
[space] if stable
[?] if unstable
[^] if high limit is out of range
[v] if low limit is out of range
[space] for positive values
[-] for negative values
9 characters with right justification
3 characters with left justification
command understood and in progress
13
14 - 16
- 81 -
)
17
18

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

As r2Ps r1Ps r2Ps r2.hApp r

Tabla de contenido