Steam Sterilization
Warning: Drying time depends on several
variables, including altitude, humidity, type of
wrap, preconditioning, size of chamber, mass of
load, material of load, and placement in chamber.
Users must verify that drying time set in their
autoclave yields dry surgical equipment.
Wrapping
Temperature
Cycle Time
Dry Time
Wrapping
Temperature
Cycle Time
Dry Time
Storage
Store packaged and sterilized instruments in a
dry, clean, and dust-free environment at room
temperatures (2 – 25°C).
Cleaning was validated using Enzol® at 1 oz/gal. at
1
35 – 40°C
Cleaning was validated using reverse osmosis
2
(RO/DI) water.
Cleaning was validated using Renu-Klenz™ at
3
1/4 oz/gal. at 35 – 40°C.
Expected Lifetime
Electrosurgical instruments will last two years,
assuming one use per day (equivalent to approximately
520 uses).
Symbols
Authorized representative in the European Community
Manufacturer
Date of manufacture
Catalog number
Batch Code
Quantity
Consult instructions for use
Caution
Made in Germany
Conformance to 93/42/EEC
Dangerous Voltage
121°C (250°F)
Pre-
Vacuum
Double
132°C
(270°F)
4 Minutes
30 Minutes 30 Minutes
Gravity
Double
50 Minutes
35 Minutes
Pre-
Vacuum
Double
134°C
(274°F)
3 Minutes
"Flash"
Pre-
Vacuum
-
132°C
(270°F)
4 Minutes
-