Fluid Filter Bag
CAUTION
Priming
WARNING
Thermo Scientific
The reservoir has a fluid bag filter designed to prevent the introduction of
particulates into the system.
Install the filter bag before starting the chiller.
Place your fingers below the front of the housing and push up on the
housing to remove it.
Fluid Bag Filter
Funnel Housing
Figure 3-6 Fluid Filter Bag
If able, pre-fill the process fluid lines. The chiller is designed to shut down if
not properly primed.
Ensure that there is enough fluid prepared to fill both the chiller and your
application. If able, pre-fill the process fluid lines to reduce the setup time.
Fill the reservoir to the max fill line on the reservoir sight tube. To prevent
the introduction of particulates into the system, fill the chiller with the
reservoir bag filter in place.
Start the chiller using the power button
As the pump fills your application add fluid to the reservoir to maintain the
fluid level.
Repeat this process until the fluid level no longer drops in the reservoir.
If you need to pause priming use the power button to turn the chiller off.
Not completely filling the chiller and process fluid lines could damage
the chiller's pump.
Install the bag.
Replace the housing. Slide its back edge
under the lip of the chiller's top panel
and then press down until the housing
snaps into place.
on the control panel.
Section 3
ThermoFlex