Electrical
Requirements
Thermo Scientific
The chiller's construction provides protection against the risk of
electrical shock by grounding appropriate metal parts. The protection
will not function unless the power cord is connected to a properly
grounded outlet. It is the user's responsibility to assure a proper
ground connection is provided.
The chiller must be installed in accordance with the National Electrical Code
and the with reference to the information on the chiller's nameplate located
on the rear.
Locate the chiller so it is near, and has easy access to, its disconnecting device.
The user is responsible to ensure that the line cord provided meets local
electrical codes. If not, contact qualified installation personnel.
The chiller is intended for use on a dedicated outlet. The ThermoFlex has
an internal circuit protection that is equivalent (approximately) to the branch
circuit rating. This is to protect the ThermoFlex, and is not intended as a
substitute for branch circuit protection.
Electrical Service Requirements (Standard temperature chillers):
ThermoFlex900
Voltage ±10%
ThermoFlex1400
Voltage ±10%
ThermoFlex2500
Voltage ±10%
200 VAC P 1, P 2 Pump
208-230 VAC P 1, P 2 Pump
200 VAC T 1 Pump
208-230 VAC T 1 Pump
* Refer to Appendix A for country specific ratings.
Frequency
100 VAC
50 Hz
115 VAC
60 Hz
200 VAC
50 Hz
208-230 VAC
60 Hz
230 VAC
50 Hz
Frequency
100 VAC
50 Hz
115 VAC
60 Hz
200 VAC
50 Hz
208-230 VAC
60 Hz
230 VAC
50 Hz
Frequency
50 Hz
60 Hz
50 Hz
60 Hz
230 VAC
50 Hz
Phase
Branch Circuit
Requirements
1Ø
1Ø
1Ø
1Ø
1Ø
*16A
, 15A
1
Phase
Branch Circuit
Requirements
1Ø
1Ø
1Ø
1Ø
1Ø
*16A
, 15A
1
Phase
Branch Circuit
Requirements
1Ø
1Ø
1Ø
1Ø
1Ø
*16A
, 15A
1
Section 3
Line Cord
Plug
15A
5-15P
15A
5-15P
15A
6-15P
15A
6-15P
, 13A
-
2
3
Line Cord
Plug
20A
5-20P
20A
5-20P
15A
6-15P
15A
6-15P
, 13A
-
2
3
Line Cord
Plug
15A
6-15P
15A
6-15P
20A
6-20P
20A
6-20P
, 13A
-
2
3
Continued on next page.
ThermoFlex
3-3