Daily Start Up
CAUTION
TEMPERATURE LIMITS
Thermo Scientific
Before starting the chiller, double check all electrical and plumbing connec-
tions. Have extra recirculating fluid on hand. If the chiller will not start refer
to Section 7 Troubleshooting.
Do not run the chiller until fluid is added.
Ensure the chiller's castors are locked.
• Ensure the circuit protector located on the rear of the chiller is in the
up position.
• Press
will automatically start.
The controller automatically sequences through the following displays:
STATUS DETECTING
DAILY STARTUP
HIGH xx.x°C
LOW xx.x°C
If Level % is not displayed on the bottom of the screen refer to Status Dis-
plays in this section.
If the fluid level is above a the required minimum value, see next page, the
pump and compressor will automatically start but only after a 30 second
delay.
If the fluid level is at or below the value, see next page, fluid will need to be
added before the compressor will start, see Initial Fill Mode on next page.
After start up, check the plumbing connections for leaks.
on the controller, the pump and the refrigeration system
xxx%
FLUID LEVEL
xx gpm
FLUID FLOW RATE
Optional Display
SETPOINT
+24.2 C
Pressure: xxx.x psi
Process Fluid Temperature Display
Figure 4-2 Daily Start Up Displays
MENU
In
Section 4
xx.x°C
SETPOINT
xx psi
PRESSURE