TEMPERATURE
TEMPERATURE is used to view/change the chiller's high and low temperature fault and warning set-
tings. If the chiller exceeds the fault setting it will shut down, the controller will display a fault message
and, if enabled, sound the alarm. If the chiller exceeds the warning setting the chiller will continue to
run, the controller will display a warning message and, if enabled, sound the alarm.
If a change is needed, press the arrows to highlight the desired setting and then press the enter button.
The highlight will flash. Press the up or down arrow to change the value. Once the desired setting is
displayed press enter to accept the change and stop the flashing.
Note You cannot set the High Warning value higher than the High Fault value. You cannot set the
Low Warning value lower than the Low Fault value.
When the desired changes are made press the left arrow, or esc button, to return to the SETTINGS
display.
TEMPERATURE
High Fault
High Warning
Low Warning
Low Fault
Thermo Scientific
High Fault/Warning Range:
+2°C to +43°C for standard temperature chillers
xxC
+2°C to +93°C for high temperature chillers
xxC
High Fault/Warning Default:
xxC
+42°C for standard temperature chillers
+92°C for high temperature chillers
xxC
Low Fault/Warning Range:
+2°C to +43°C for all chillers
Low Fault/Warning Default:
+3°C for all chillers
Section 4
ThermoFlex 4-13