Condenser Filter

CAUTION
WARNING
WARNING
Thermo Scientific
Failure to clean/replace the condenser filter will cause a loss of
cooling capacity and lead to premature failure of the cooling
system.
ThermoFlex900s - 5000s
Clean the filter through the grill using a vacuum with a soft-bristle brush.
When it is time for a more thorough cleaning, remove the one-piece grill
assembly by first pulling the bottom of the assembly away from the chiller
and then pulling it away from the top.
The condenser framing and fins located behind the grill assembly are
very sharp. Use caution when removing the assembly.
Note ThermoFlex900s - 5000s water-cooled chillers have an embedded
screw(s) located at the top (and bottom) of the grill securing it to the chiller.
Loosen the screw(s) to remove the grill.
Water-cooled chillers also have a fan with sharp blades, ensure the
chiller is off before removing the assembly.
Shake off as much of the excess water as possible before reinstalling. Press
the grill back into place.
For water-cooled chillers, tighten the screw(s) at the top (and bottom) of
the grill.
Replacement grill assemblies are available from Thermo Fisher.
Screw
(Water-cooled chillers)
Figure 6-3 ThermoFlex900s - 5000s Condenser Grill Removal
ThermoFlex3500s - 5000s water-cooled
chillers have two screws securing the
condenser. One screw on the top and one on
the bottom.
CAUTION - Sharp edges