Appendix D
USER RANGE SETTINGS
This menu allows the you to customize the scaling of the analog input and DAC output to meet the application
USER RANGE SETTINGS
needs. With
The ThermoFlex supports three standard analog interface types and defaults to the following scaling:
Voltage
MilliVolts
Current
D-4
ThermoFlex
highlighted press enter to display:
USER RANGE SETTINGS
Voltage
Millivolts
Current
0v – 10v = the operating range of the chiller
(5° – 40° in standard chillers, 5° – 90° in high temp chillers)
10mv/°C
4ma – 20ma = the operating range of the chiller
Thermo Scientific