Section 3

Fluid Top Off

Ensure the reservoir cap is at a safe handling temperature before
removing.
Remove the reservoir cap by unscrewing it counterclockwise.
To prevent the introduction of particulates into the system, fill the chiller
with the reservoir bag filter in place.
The reservoir has a sight tube and ball for easy fluid level monitoring.
Slowly fill the reservoir with clean process fluid through the funnel only,
failure to comply may result in internal spillage.
Note Filling the reservoir above MAX LeVeL fill line will result in a over
flow error (Over Flow) causing the chiller to shut down. Also, fluids
expand when heated.
Note Adding fluid that has a temperature differential with the fluid
already in the reservoir will temporarily affect the chiller's stability
performance.
Before replacing the reservoir cap ensure the reservoir sight tube
CAUTION
ball stopper is securely in place.
Reservoir Ball Stopper
Figure 3-9 Reservoir Ball Stopper