Section 4
SETPOINT
Thermo Scientific
Note Depressing and holding the enter button for two seconds before
starting the chiller allows you to view, as well as make changes to, the set-
point.
Press the arrows to highlight SETPOINT.
SETPOINT
+24.2 C
Pressure: xxx.x psi
Press enter and the setpoint value appears. If needed, press enter again and
the value starts flashing indicating the setpoint can be changed.
The setpoint range for standard temperature chillers is +5°C to +40°C,
the range for high temperature chillers is +5°C to +90°C. Press the up and
down arrows to change the setpoint to the desired value.
Once the desired value is displayed press the enter button again to
accept the change and to stop the flashing.
Return to the Process Fluid Temperature Display by pressing the left
arrow or esc button.
SETPOINT
+24.2 C
Pressure: xxx.x psi
MENU
In
xx.xC
MENU
In
ThermoFlex 4-6