Live Data Stream; Driver Installation - Hutchinson Technology InSpectra StO2 650 Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

Data Transfer
(continued)

Live Data Stream

The InSpectra
StO
System Measurement can be obtained
2
through the RS232 serial data port
InSpectra
StO
Monitor.
2
When the InSpectra
StO
measurement data is sent out both ports every two (2) seconds.
the format for the data stream is:
Date (YYYY-MM-DD) Time (hh:mm:ss) Valid Reading Flag
(1 = Valid)
% StO2
Example of data output: 2005-04-21, 9:46:01, 1, 78, 9.2
Explanation: This reading was taken on April 21, 2005 at 9:46:01; the reading was valid, with 78% StO
and THI of 9.2.
Data can be collected using a PC running a terminal program. When connecting to the monitor, use the
following settings:
• Baud Rate: 115200
• Data Bits: 8
Stored data collected during the last 96 hours can be retrieved from the monitor by sending the ASCII "xfer"
command, followed by a carriage return, from the PC. The monitor will stream all stored data in the format above.
The transfer data stream may be stopped by sending the "xfer 0" command.

Driver Installation

The proper FTDI Driver must be installed on the PC collecting data to use the USB port. The VCP driver can be
downloaded from:
The USB integrated circuit manufacturer's website: http://www.ftdichip.com/FTDrivers.htm.
NotE: You can download a royalty-free USB driver compatible with your operating system and the FT232BM
Integrated Circuit installed in the InSpectra StO
Verify proper operation before clinical use when connecting the InSpectra StO
Monitor to an external IT device.
cautIoN: Any external IT device connected to this monitor must be certified according to IEC
Standard 950.
• C ompliance with IEC Standard 60601-1-1 system requirements is required for all combinations of equipment
connected to this monitor.
• A ny person connecting peripheral equipment to the data output port is configuring a medical system and is
therefore responsible for compliance to system standard IEC Standard 60601-1-1 and the electromagnetic
compatibility system standard IEC Standard 60601-1-2.
O
WarNING: Possible Electrical Shock Hazard. Do not simultaneously touch a patient and an external
device, such as a computer, that is connected to the InSpectra StO
the USB port or the RS232 port.
For additional information on capturing data externally, contact your Hutchinson Technology representative
or the Customer Service Center.
u.S. and canada Service center
Telephone: 800.419.1007 or 320.587.1926
Fax: 320.587.1555
Email: biom.usa@hti.htch.com
24
Data Transfer
on the back of the
1
Sensor is connected to the monitor,
2
THI
• Parity Bits: None
• Stop Bits: 1
• Flow Control: None
Tissue Oxygenation Monitor. Instructions are on the website.
2
2
Tissue Oxygenation Monitor through
2
European Service center
Telephone: +31 26 365 33 70
Fax: +31 26 365 33 72
Email: biom.eu@hti.htch.com
1
2
figure 24.1
2
Tissue Oxygenation

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido