Recommended Separation Distances Between Portable And Mobile; Rf Communications Equipment And The Inspectra - Hutchinson Technology InSpectra StO2 650 Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

Technical Addendum
recommended Separation Distances Between portable and Mobile rF
communications Equipment and the InSpectra
The InSpectra StO
Tissue Oxygenation Monitor is intended for use in an electromagnetic environment in which
2
radiated RF disturbances are controlled. The customer or the user of the InSpectra StO
Monitor can help prevent electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and the InSpectra StO
recommended below, according to the maximum output power of the communications equipment.
rated maximum
output power of
150kHz to 80MHz
transmitter
(W)
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance in meters
(m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum
output power rating of the tranmitter in watts (W) according to the transmitter manufacturer.
NotE: At 80 MHz and 800 MHz, the higher frequency range applies.
NotE: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
Separation distance according to frequency of transmitter (m)
80MHz to 800MHz
d = 1,2√P
d = 1,2√P
0,12
0,38
1,2
3,8
12
Sto
tissue oxygenation Monitor
2
Tissue Oxygenation Monitor as
2
0,12
0,38
1,2
3,8
12
Tissue Oxygenation
2
800MHz to 2,5GHz
d = 2,3√P
0,23
0,73
2,3
7,3
23
table 203
37
Technical Addendum