Specifications (continued)
Flyte Helmet REF 0408-600-000, 0408-635-000 and 0408-645-000 are intended for use in the electromagnetic environment specified below. The customer or the
user of Flyte Helmet REF 0408-600-000, 0408-635-000 and 0408-645-000 should assure that it is used in such an environment.
NOTE: The values provided in the table below have changed due to 60601-1-2 4th Edition requirements.
Immunity test
Electrostatic discharge (ESD)
IEC 61000-4-2
Electrical fast transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short interruptions and
voltage variations on power supply
input lines
IEC 61000-4-11
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
Recommended separation distances between portable and mobile RF equipment and the Flyte Helmet REF 0408-600-000, 0408-635-000 and 0408-645-000
Flyte Helmet REF 0408-600-000, 0408-635-000 and 0408-645-000 are intended for use in the electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of Flyte Helmet REF 0408-600-000, 0408-635-000 and 0408-645-000 can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF equipment (transmitters) and Flyte Helmet REF 0408-600-000, 0408-635-000 and 0408-645-000
Rated maximum output power of
transmitter
W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation
applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
10
Guidance and manufacturer's declaration - electromagnetic immunity
IEC 60601 test level
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air
± 2 kV at 100 kHz repetition frequency
for power supply lines
± 1 kV at 100 kHz repetition frequency
for input/output lines
± 0.5 kV, ± 1 kV line(s) to line(s)
± 0.5 kV, ± 1 kV, ± 2 kV line(s) to earth
<5% U
(>95% dip in U
)
T
T
for 0.5 cycle
40% U
(60% dip in U
)
T
T
for 5 cycles
0% U
(100% dip in U
)
T
T
for 0.5 cycle at 0°, 45°, 90°, 135°, 180°,
225°, 270°, and 315°
0% U
(100% dip in U
)
T
T
for 1 cycle at 0°
70% U
(30% dip in U
)
T
T
for 25 and 30 cycles at 0°
<5% U
(>95% dip in U
)
T
T
for 5 s
0% U
(100% dip in U
)
T
T
for 250/300 cycles
3 A/m, 30 A/m at 50 and 60 Hz
IEC 60601-1-2 3rd Edition:
as recommended below, according to the maximum output power of the equipment.
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d=1.2√P
0.12
0.38
1.2
3.8
12
Compliance level
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air
Not Applicable
Not Applicable
Not Applicable
3 A/m, 30 A/m at 50 and 60 Hz
m
80 MHz to 800 MHz
d=1.2√P
0.12
0.38
1.2
3.8
12
0408-600-705 Rev-AB
Electromagnetic environment -
guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Not Applicable
Not Applicable
Not Applicable
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commercial
or hospital environment.
800 MHz to 2.5 GHz
d=2.3√P
0.23
0.74
2.3
7.4
23