Changing Prosthesis Settings - Ottobock Genium 3B1-2 Instrucciones De Uso

Ocultar thumbs Ver también para Genium 3B1-2:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 187
→ A beep and vibration signal will sound to confirm that the movement pattern has been
recognised.
INFORMATION: If this beep and vibration signal does not sound, the requirements
were not observed when bouncing the foot or mute mode (silent mode) is activ­
ated. For more information about mute mode, see the section "Mute mode" (see
Page 77).
4) After the beep and vibration signal sounds, keep the prosthetic leg extended and still for 1
second.
→ A confirmation signal will sound to indicate that the prosthesis has successfully switched over
to basic mode.
INFORMATION: If this confirmation signal does not sound, the leg with the prosthesis
was not correctly kept still or mute mode (silent mode) is activated. Repeat the pro­
cess to correctly switch to the required mode. For more information about mute mode,
see the section "Mute mode" (see Page 77).

8.3 Changing prosthesis settings

Once an active connection to a prosthesis has been established, the settings of the mode cur­
rently active can be changed using the cockpit app or the remote control (optional accessory).
For information on changing the prosthesis settings with the remote control (optional accessory),
see the remote control instructions for use.
INFORMATION
Bluetooth on the prosthesis must be switched on to change the prosthesis settings.
If Bluetooth is switched off, it can be turned on by turning the prosthesis upside-down or by
connecting/disconnecting the battery charger. Bluetooth is then turned on for approx. 2
minutes. The connection must be established during this period.
Information for changing the prosthesis settings
Before changing settings, always check the main menu of the cockpit app or the display of
the remote control (optional accessory) to make sure the correct prosthesis has been selec­
ted. Otherwise, parameters could be changed for the wrong prosthesis.
It is not possible to change prosthesis settings nor to switch to a different mode while the
prosthesis battery is being charged. Only the status of the prosthesis can be called up.
Instead of the
pit app.
The prosthetist's setting is located in the middle of the scale. After making adjustments, this
setting can be restored by tapping the "Standard" button (Cockpit app) or by moving the
slider control into the middle (remote control).
Prosthesis settings should be optimised using the adjustment software. The cockpit app or
remote control (optional accessory) is not intended for use by the prosthetist in order to set up
the prosthesis. The behaviour of the prosthesis can be changed to a certain extent on a day-
to-day basis using the remote control or app (e.g. while becoming accustomed to the pros­
thesis). The prosthetist can use the adjustment software to trace these changes at the next
appointment.
If the settings of a MyMode are to be modified, one must first switch to this MyMode.
 symbol, the
 symbol appears in the bottom row of the screen in the cock­
Ottobock | 73

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Genium 3b1-2-st

Tabla de contenido