Error message
14.2.4 Status signals
Battery charger connected
LED on
Status
power
LED on
supply
battery
charger
Battery charger disconnected
Beep
Vibration
signal
signal
1x short
1x short
88 | Ottobock
Cause
A current connection to
the prosthesis was inter
rupted.
Event
Power supply and battery charger operational. Battery charger not yet
connected to receiver.
Battery charger connected to receiver, good connection.
This display turns off automatically after approximately one minute to
avoid bothersome light at night. The charging process is not affected.
Event
Self-test completed successfully.
Product is operational.
Correction
Check the following points:
•
Distance from the prosthesis to the
device
•
Battery charge level of the prosthesis
•
Bluetooth on the prosthesis switched
on? (see Page 76)
•
If multiple prostheses were stored in
the remote control, was the correct
prosthesis selected?
Resolution