Maintenance and Cleaning

11.1. Routine Maintenance

WARNING: Do not use lubricants on this device or any of its accessories.
CAUTION: Replace the cannula on a regular basis. Check with your distributor or clinician to
determine how often the cannula should be replaced.
Device will indicate with an alarm when service is required. (Also, see Chapter 10 - Trouble-
shooting.)

11.2. Cleaning and Disinfection

WARNING: Do not submerge this device in liquid. Do not expose to water or precipitation.
Do not expose to dusty conditions.
CAUTION: Do not use cleaning agents other than those specified in this manual. Allow the
cleaning solution to dry from the cleaned surface before use.
CAUTION: Always disconnect power and turn off this device before cleaning.
Clean the exterior with a soft cloth slightly dampened with soapy water or with anti-bacterial
wipes (Isopropyl alcohol 70% solution).
For disinfecting, use a MadaCide-FDW-Plus wipe or equivalent and follow the manufacturer's
instructions. (Manufacturer - Mada Medical Products Inc., www.madamedical.com)
Important: The device should receive an external cleaning weekly, accessories should be cleaned
as needed. The device exterior and accessories should be cleaned and disinfected and the patient
filter replaced prior to delivering to a new patient.
Nasal cannula: Refer to the original manufacturer's instructions for cleaning the nasal can-
nula.

11.3. Service Life

The expected service life of the device is 5 years, except for the sieve modules. The service
life of the sieve modules will depend on the operating conditions. The sieve modules are an
internal component of the device and should only be replaced by a trained person. If the in-
take and exhaust vents are not blocked and the check vents alarm is activated contact your
distributor.
27/292
EN