Ventricular capture control
Parameter
Range of values
A/RV/LV capture control ATM (monitoring only)
ON; OFF
Min. amplitude
0.7 V
Threshold test start
2.4; 3.0; 3.6; 4.2; 4.8 V
RV safety margin
0.3 ... (0.1) ... 1.2 V
LV safety margin
1.0; 1.2 V
Search type
Interval; time of day
Interval
0.1; 0.3; 1; 3; 6; 12; 24 h
Time of day
00:00 ... (10 min) ... 23:50
Atrial overdrive pacing
Parameter
Range of values
Atrial overdrive pacing
ON; OFF
Lead configuration
Parameter
Range of values
Sensing polarity A and RV/LV
Unipolar; bipolar
Pacing polarity A and RV
Unipolar; bipolar
Pacing polarity LV
LV tip -> LV ring
LV tip -> RV ring
LV ring -> LV tip
LV ring -> RV ring
LV tip -> housing
LV ring -> housing
en • English
IEGM recordings
Standard
Parameter
ON
IEGM recordings
Types of IEGM recordings
0.7 V
3.0 V
1.0 V
1.0 V
IEGM recording prior to event
Time of day
IEGM signal
24 h
Rates for statistics
2:00 hh:mm
Parameter
High atrial rate (HAR)
Standard
OFF
High ventricular rate (HVR)
HVR counter
Standard
Rate Adaptation
Unipolar
Rate adaptation via Closed Loop Stimulation
Unipolar
CLS modes:
LV tip –> housing
Parameter
Max. CLS rate
CLS response
Resting rate control
Vp required
17
Range of values
Standard
20 (quantity); each max. 10 s
High atrial rate (HAR)
HAR
Mode switching
High ventricular rate (HVR)
ON
Patient-triggered
OFF
0; 25; 50; 75; 100%
75%
Filtered; unfiltered
Filtered
Range of values
Standard
100 ... (5) ... 250 bpm
200 bpm
600; 572 ... 245; 240 ms
300 ms
150 ... (5) ... 200 bpm
180 bpm
400; 387 ... 308; 300 ms
333 ms
4; 8; 12; 16
8
Range of values
80 ... (5) ... 160 bpm
Very low; low; medium; high; very high
OFF; +10 ... (10) ... +50 bpm
Yes; no
Standard
120 bpm
Medium
+20 bpm
Yes