6.2 Connect battery charger to the product
INFORMATION
Do not move the knee joint while it conducts the self-test immediately after disconnecting the
charger. Otherwise, an error may occur; if this happens, the problem can be corrected by
reconnecting and then disconnecting the charger.
INFORMATION
To make the operating time of the prosthesis as long as possible, the charger should not be
removed until immediately before the prosthesis is used.
Indication of the charging process:
Charger
Battery is charging.
The on time of the LED indicates the current battery charge level.
The on time of the LED gets longer as the charge level increases. It only flashes
briefly at the start of the charging process and stays on continuously at the end of
the charging process.
Battery is fully charged, or the temperature fell above/below the permissible range
during charging. Check current charge level (see Page 50).
1) Remove the prosthesis.
2) Connect the inductive charger to the receiver of the charging unit
on the back of the product.
Make sure the contact surfaces are clean, with no objects adhering
to them.
→ The charger is held in place by a magnet.
→ A correct connection between the battery charger and the
product is indicated by feedback (see Page 65).
3) The charging process starts.
→ Once the product battery is fully charged, the LED on the bat
tery charger lights up green.
4) After the charging process is complete, remove the inductive char
ger from the receiver and hold the product still.
→ A self-test is performed, and the product should not be moved
while this is in progress. The joint is ready for operation only
after corresponding feedback (see Page 65).
5) Put the prosthesis on.
49