Therefore, keeping the following minimum distances to these HF communication devices is
recommended:
• Mobile phone GSM 850/GSM 900: 0.50 m
• Mobile phone GSM 1800/GSM 1900/UMTS: 0.35 m
• DECT cordless phones incl. base station: 0.18 m
• WiFi (routers, access points,...): 0.11 m
• Bluetooth devices (third-party products not approved by Ottobock): 0.11 m
CAUTION
Wearing in extreme ambient temperatures
Injury due to faulty control or malfunction of the Axon-Bus prosthethic system.
Avoid wearing the prosthesis in areas with temperatures outside of the permissible range.
See the section "Technical Data".

4.5 Information on Use

CAUTION
Using the prosthesis with pointed or sharp objects (e.g. knives in the kitchen)
Injury due to unintentional movements.
Use extra caution when using the prosthesis for handling pointed or sharp objects.
CAUTION
Excessive strain or overloading due to unusual activities
Injury due to faulty control or malfunction of the Axon-Bus prosthethic system.
The Axon-Bus prosthethic system was developed for everyday use and must not be used for
unusual activities. These unusual activities include, for example, sports with excessive strain
and/or shocks to the wrist unit (pushups, downhill, mountain biking) or extreme sports (free
climbing, paragliding, etc.).
Careful handling of the Axon-Bus prosthethic system and the Axon-Bus components not only
increases their service life, but is primarily in the interest of your own safety.
If the Axon-Bus prosthethic system and the Axon-Bus components have been subjected to
extreme loads (e.g. through a fall or similar occurrence) the Axon-Bus prosthethic system
must be inspected by a prosthetist immediately for possible damage. The contact person for
this is the responsible prosthetist, who will send the Axon-Bus prosthethic system to
Ottobock Service if necessary.
CAUTION
Mechanical overload caused by mechanical influences or external loads
Injury due to faulty control or malfunction of the Axon-Bus prosthethic system.
Do not subject the Axon-Bus components to mechanical vibrations or impacts.
Check the Axon-Bus components for any visible damage before each use.
CAUTION
Penetration of dirt and moisture
Injury due to faulty control or malfunction of the Axon-Bus prosthethic system.
Ensure that neither solid particles nor liquids can penetrate into the Axon-Bus prosthethic
system or Axon-Bus components (e.g. the Axon-Bus gripping component).