Legal Information - Ottobock Axon-Bus Prosthetic System Instrucciones De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 83
Action/Event
Outside operating temper­
ature range
Action/Event
Battery not charging
Battery not charging
Error message
Service message
Error message
12 Disposal
This product may not be disposed of with regular domestic waste in all jurisdictions. Dis­
posal that does not comply with the regulations of your country may have a detrimental
impact on health and the environment. Please observe the information provided by the
responsible authorities in your country regarding return and collection processes.

13 Legal Information

13.1 Liability
Otto Bock Healthcare Products GmbH, hereafter referred to as manufacturer, assumes liability
only if the user complies with the processing, operating and maintenance instructions as well as
the service intervals. The manufacturer explicitly states that this device may only be used in com­
bination with components that were authorised by the manufacturer (see instructions for use and
catalogues). The manufacturer does not assume liability for damage caused by component com­
binations which it did not authorise.
The device may only be opened and repaired by authorised Ottobock technicians.
Charging
Beeper
receptacle
 
 
Battery charger
Probable cause
 
 
The charging
plug is not prop­
erly connected to
the charging
receptacle.
The charging
plug or charging
receptacle con­
tacts are dirty.
 
 
The battery or
   
battery charger
 
 
temperature is
outside the per­
missible range.
 
 
There is an error
in the Axon-Bus
prosthetic sys­
tem.
 
 
Service required. Contact your prosthetist.
 
 
The battery char­
ger or the power
supply unit is
defective.
Remedy
Allow the Axon-Bus pros­
thetic system to warm up or
cool down; see the section
'Technical Data' (see
Page 41).
Remedy
Disconnect and then
reconnect the charging
plug.
Use a cotton swab and a
mild soap solution to clean
the contacts of both com­
ponents.
Disconnect the charging
plug and allow the battery
or the charger to warm up
or cool down.
Contact your prosthetist.
Contact your prosthetist.
Ottobock | 39

Publicidad

Tabla de contenido
loading

Tabla de contenido