Shift Detected; Warning - Ottobock 3D L.A.S.A.R. Posture Instrucciones De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 257
Use

5.4.5 Shift detected

Shift detected
Platform shift detected.
0.0 Nm
0.0 %
Shift detected
Platform shift detected.
Cancel

5.4.6 Warning

Warning
Measuring accuracy too low.
Realign the measurement plate.
0.0 Nm
0.0 %
Warning
Measuring accuracy too low.
Realign the measurement plate.
Cancel
100
Live view
Platform shift. Tap to recalibrate...
Cancel
Recalibrate
0.0 Nm
0.0 %
1
2
Recalibrate
Live view
Calibration failed. Tap to recalibrate...
Cancel
Recalibrate
0.0 Nm
0.0 %
1
2
Recalibrate
Error message text:
: "Shift detected"
: "Platform shift detected"
: "Cancel"
: "Recalibrate"
Possible causes:
The camera stand or force measurement plate was
moved
All LEDs were pushed out of position
Underexposed image (black)
Measures:
► Carry out the appropriate measures depending on
the cause (for example align the force measurement
plate and camera stands so that all LEDs are recog­
nised within the outer blue frame; restart the mini-
computer in case of incorrect exposure).
► 1x tap on "Recalibrate".
Error message text:
: "Warning"
: "Measuring accuracy too low."
: "Realign the measurement plate." – Position the
force measurement plate closer to the camera stand to
use the full camera resolution.
: "Cancel"
: "Recalibrate"
Possible causes:
Force measurement plate rotated too much relative
to the camera
Camera stand tilted away from the force measure­
ment plate too much
Camera too far away from the force measurement
plate
Measures:
► Carry out the appropriate measures depending on
the cause (for example align the force measurement
plate and camera stands so that all LEDs are recog­
nised within the outer blue frame, that the
crosshairs of the inner blue frame are centred over
the middle of the force measurement plate, and that
the hinges on the force measurement plate are
aligned with one of the two centre lines of the
crosshairs; align the camera at a 90° angle; reduce
the distance between the camera and force meas­
urement plate).
► 1x tap on "Recalibrate".
743L500 3D L.A.S.A.R. Posture